Nuclear Magnetic Resonance Spectroscopy Buying Guide

Nuclear Magnetic Resonance (NMR) spectroscopy identifies molecular structure by measuring radio-frequency absorption of atomic nuclei in a strong magnetic field. The decisive trade-off is field strength versus cost and siting: higher tesla delivers better resolution and sensitivity but demands cryogens, cryocoolers, and reinforced floors. Buy a superconducting FT-NMR only when your chemistry requires <0.1 ppm resolution; choose benchtop permanent-magnet or time-domain systems when you need routine QA or relaxometry without cryogen logistics.

Field Strength:1-2 T benchtop / 7-11.7 T superconductingMagnet Type:Permanent / Superconducting solenoidProbe Configuration:5 mm BB / inverse / flow / 10 mmCryogen Requirement:He + N2 refills / cryocooler / none

Key takeaways

  • Field strength tiers: benchtop permanent magnets at 42-80 MHz (1-2 T) vs superconducting solenoids at 300-500 MHz (7-11.7 T) define resolution and sensitivity ceilings.
  • Probe architecture drives application fit: 5 mm broadband switchable probes for 1H/13C, inverse cryoprobes for trace-mass sensitivity, and 10 mm or flow probes for high-throughput screening.
  • Superconducting magnets require liquid helium (roughly every 6-12 months) and liquid nitrogen (every 2-4 weeks) unless a cryogen-free cryocooler is specified.

At a glance

Price range
Varies by listing
Category
Spectroscopy, Analysis & Detection Instruments
Last updated
Nov 28, 2026

Current offers

No live listings right now. Search directly on:

Highlights

  • Field strength tiers: benchtop permanent magnets at 42-80 MHz (1-2 T) vs superconducting solenoids at 300-500 MHz (7-11.7 T) define resolution and sensitivity ceilings.
  • Probe architecture drives application fit: 5 mm broadband switchable probes for 1H/13C, inverse cryoprobes for trace-mass sensitivity, and 10 mm or flow probes for high-throughput screening.
  • Superconducting magnets require liquid helium (roughly every 6-12 months) and liquid nitrogen (every 2-4 weeks) unless a cryogen-free cryocooler is specified.
  • Time-domain (TD-NMR) benchtop analyzers at 0.5 T measure relaxation and solid-fat content without chemical-shift resolution, at a fraction of FT-NMR cost.

Best use cases

  • Best fit: synthetic organic chemistry and structure elucidation — choose a 400 MHz superconducting FT-NMR with a 5 mm broadband probe and automated sample changer.
  • Best fit: routine food, polymer, or agricultural QC — choose a 60-80 MHz benchtop permanent-magnet FT-NMR or TD-NMR analyzer; skip cryogen-dependent systems.
  • Wrong fit: trace impurity quantification below 0.01% — standard 5 mm probes lack sensitivity; specify an inverse cryoprobe or switch to LC-MS/MS.
  • Wrong fit: a shared multi-user lab with no helium delivery contract or floor-load survey — defer superconducting purchase until cryogen logistics and structural reinforcement are confirmed.

Buying tips

  • Match field strength to nucleus: 1H-only screening works at 60 MHz; 13C and 2D experiments need 300 MHz or higher and a broadband probe.
  • Confirm magnet stray-field footprint: a 400 MHz superconducting magnet needs roughly 2.0-2.5 m radial 5-gauss clearance, often more; benchtop units need under 1 m.
  • Specify probe and automation upfront — probe, sample changer, and software can add 30-50% beyond the base magnet and console cost.
  • Require a cryogen service contract and helium recovery system for any superconducting magnet; without recovery, helium boil-off dominates annual operating cost.

Specs

Field Strength1-2 T benchtop / 7-11.7 T superconducting
Magnet TypePermanent / Superconducting solenoid
Probe Configuration5 mm BB / inverse / flow / 10 mm
Cryogen RequirementHe + N2 refills / cryocooler / none

Related searches

What to check before buying

  • Floor loading: superconducting magnet plus console typically exceeds 500 kg concentrated load; verify structural slab capacity and vibration isolation.
  • Cryogen supply: confirm liquid helium and nitrogen delivery within 48 hours and on-site storage/dewar capacity before committing.
  • Site environment: stable temperature within ±1 °C and humidity below 60% RH; benchtop permanent magnets are sensitive to temperature drift.
  • Power and networking: dedicated 208-240 V circuit with UPS for the console; wired Ethernet for instrument control and LIMS integration.