High Voltage Scope Probe Buying Guide
A high voltage scope probe is a compensated attenuator that scales dangerous potentials down to a scope's front end while preserving signal fidelity. The core trade-off is voltage rating versus bandwidth and input impedance: higher voltage probes use larger series resistors and longer compensation networks that sap bandwidth and load the circuit more. Choose a probe rated at least 1.5x your maximum working voltage with a bandwidth at or above your signal's highest significant harmonic, and reject generic 'HV' probes with no published derating curve or safety certification.
Key takeaways
- Voltage derating is frequency-dependent: a 1 kV DC / 1 kV peak AC probe typically falls to 500–600 V at 1 MHz and far lower at 10 MHz, so always read the V-vs-frequency curve.
- 100:1 attenuation is the high-voltage standard; 10:1 probes top out around 300–600 V and are not substitutes for true HV work.
- Input impedance of 100 MΩ (100:1) versus 10 MΩ (10:1) reduces circuit loading, but cable capacitance and compensation range still affect accuracy.
At a glance
- Price range
- Varies by listing
- Category
- Test Leads, Probes, Fixtures & Accessories
- Last updated
- Nov 28, 2026
Current offers
Highlights
- Voltage derating is frequency-dependent: a 1 kV DC / 1 kV peak AC probe typically falls to 500–600 V at 1 MHz and far lower at 10 MHz, so always read the V-vs-frequency curve.
- 100:1 attenuation is the high-voltage standard; 10:1 probes top out around 300–600 V and are not substitutes for true HV work.
- Input impedance of 100 MΩ (100:1) versus 10 MΩ (10:1) reduces circuit loading, but cable capacitance and compensation range still affect accuracy.
- Safety ratings matter: CAT II/III ratings and IEC 61010 certification separate legitimate HV probes from unbranded units with only a marketing voltage number.
Best use cases
- Best fit: switch-mode power supply and inverter repair up to 1 kV — choose a 100:1 probe rated 1–2 kV with at least 20 MHz bandwidth and a CAT II/III safety rating.
- Best fit: differential measurements on floating or high-side nodes — use a true high-voltage differential probe (e.g., 1–2 kV differential, 25–100 MHz) rather than two single-ended probes.
- Wrong fit: probing 3.3 V/5 V logic or low-voltage audio — a standard 10:1, 100–500 MHz passive probe is cheaper, lower-loading, and more accurate; skip HV probes entirely.
- Wrong fit: measuring mains 230/400 V CAT III distribution panels with a hobby-grade 100:1 probe — select a probe explicitly rated for CAT III at the working voltage, or use a properly rated differential probe.
Buying tips
- Match attenuation to voltage: 10:1 probes are for ≤600 V; 100:1 for 1–2 kV; 1000:1 for multi-kV CRT/industrial work. Never exceed the probe's DC or AC peak rating.
- Check the derating curve: a probe rated 1 kV DC may only handle 300–500 V at 1 MHz; your signal's frequency content determines the real usable voltage.
- Bandwidth rule: choose a probe with bandwidth ≥ 3–5x the highest harmonic you need to see; for switching edges, that often means 20–50 MHz minimum even for 100 kHz switching.
- Compensation range must cover your scope's input capacitance (typically 10–20 pF); a probe that cannot be compensated will show rounded edges or overshoot.
Specs
| Attenuation | 10:1 / 100:1 / 1000:1 fixed or switchable |
|---|---|
| Max Voltage (DC/peak) | 300 V – 6 kV (derated by frequency) |
| Bandwidth | 20 MHz – 400 MHz by tier |
| Input Impedance | 10 MΩ (10:1) / 100 MΩ (100:1) |
Related searches
What to check before buying
- Verify your scope's maximum input voltage and input capacitance — the probe's compensation range and safety rating must match or exceed the scope's front-end limits.
- Confirm the probe's connector type (BNC standard) and whether your scope needs a 50 Ω or 1 MΩ termination; HV passive probes are designed for 1 MΩ scopes.
- Measure your maximum working voltage and frequency, then compare against the probe's derating chart — not just its headline DC rating.
- Check physical reach: probe tip length, ground lead length, and insulation rating must suit your test point spacing and environment (e.g., CAT III requires adequate creepage and cable rating).