Lauda Re 106 Buying Guide

The Lauda RE 106 is a classic refrigerated circulating thermostat (chiller) used to hold baths, reactors, and analytical instruments at precise temperatures below ambient. The decisive trade-off is cooling capacity versus temperature stability: compressor-based chillers like the RE 106 deliver strong pull-down but rely on older refrigerants and analog controls, while modern inverter units offer tighter control at higher cost. Buy the RE 106 only if you need robust sub-ambient cooling to roughly -10°C with a large bath volume; otherwise choose a newer inverter chiller for tighter stability and lower energy draw.

Cooling Type:Compressor refrigerated circulatorTemp Range:-10°C to +150°CStability:±0.1–0.3 K (bath dependent)Power Supply:230 V / 50 Hz (requires transformer in US)

Key takeaways

  • RE 106 is a compressor refrigerated circulator with roughly 1.0–1.2 kW cooling capacity at 20°C, suitable for bath volumes up to about 30–50 L.
  • Temperature range spans approximately -10°C to +150°C, but stability is typically ±0.1–0.3 K depending on bath size, fluid, and insulation.
  • Older RE-series units use R134a refrigerant and analog or early digital PID controllers; modern equivalents use R513A/R290 and inverter compressors for better efficiency.

At a glance

Price range
Varies by listing
Category
Lab Heating, Cooling & Vacuum Equipment
Last updated
Nov 28, 2026

Current offers

No live listings right now. Search directly on:

Highlights

  • RE 106 is a compressor refrigerated circulator with roughly 1.0–1.2 kW cooling capacity at 20°C, suitable for bath volumes up to about 30–50 L.
  • Temperature range spans approximately -10°C to +150°C, but stability is typically ±0.1–0.3 K depending on bath size, fluid, and insulation.
  • Older RE-series units use R134a refrigerant and analog or early digital PID controllers; modern equivalents use R513A/R290 and inverter compressors for better efficiency.
  • Pump flow is typically 10–20 L/min at low pressure; check whether your application needs higher pressure for closed-loop reactor jackets versus open bath circulation.

Best use cases

  • Best fit: batch cooling of analytical baths, viscometers, or reactors in the -10°C to +80°C range where a durable compressor chiller is preferred over a recirculating chiller with tighter ±0.01 K stability.
  • Best fit: labs with existing Lauda bath accessories and 230 V/50 Hz supply—the RE 106 integrates with standard M16x1 or clamp fittings and does not need a separate heat exchanger.
  • Wrong fit: applications requiring temperatures below -15°C or stability better than ±0.05 K—choose a modern inverter chiller or a cryostat instead.
  • Wrong fit: 120 V/60 Hz North American labs without a step-up transformer—the RE 106 is a 230 V European-market unit and requires voltage conversion.

Buying tips

  • Match cooling capacity to heat load: for a 20 L bath at 20°C ambient, a 1 kW chiller holds setpoint; add 20–30% margin if your process generates heat or the bath is poorly insulated.
  • Verify refrigerant: R134a units are being phased down in many regions; R513A or R290 models are preferable for long-term serviceability and lower GWP.
  • Check controller generation: analog dial units lack digital setpoint and alarm outputs; prefer RE 106 units with the newer LED/PID controller if you need reproducible setpoints and RS-232.
  • Confirm pump pressure: open-bath circulation needs low pressure/high flow; jacketed reactors need higher pressure—compare the pump curve, not just flow rate.

Specs

Cooling TypeCompressor refrigerated circulator
Temp Range-10°C to +150°C
Stability±0.1–0.3 K (bath dependent)
Power Supply230 V / 50 Hz (requires transformer in US)

Related searches

What to check before buying

  • Measure your bath or vessel volume and insulation; the RE 106 is best for 15–40 L—larger baths will struggle to reach -10°C quickly.
  • Confirm your facility voltage: 230 V/50 Hz is standard for the RE 106; in 120 V regions you need a 2 kVA step-up transformer with proper grounding.
  • Check available floor or bench space and clearance for rear ventilation—compressor chillers need 10–15 cm around the condenser.
  • Verify fluid compatibility: use a glycol-water mix for sub-ambient operation and confirm the pump seals are rated for your temperature range.