Surface Plasmon Resonance Spectroscopy Buying Guide
Surface plasmon resonance (SPR) spectroscopy measures real-time, label-free biomolecular binding by detecting refractive-index changes at a metal-dielectric interface. The core trade-off is between angular-scanning (high resolution, moving parts) and imaging/array (throughput, lower per-spot sensitivity) architectures, with fluidic automation and sensor-chip chemistry determining data quality. Choose a fixed-angle or scanning platform for kinetics on one channel; choose an imaging SPR only when multiplexed binding maps matter more than sub-nanomolar detection.
Key takeaways
- Kretschmann prism coupling remains the dominant architecture; angular interrogation resolves ~10^-6 RIU while intensity interrogation trades resolution for speed and low cost.
- Sensor chips use a ~50 nm gold film on glass; self-assembled alkanethiol monolayers (e.g., 11-mercaptoundecanoic acid) provide carboxyl surfaces for amine coupling of ligands.
- Flow cells range from single-channel to 4-channel or array formats; integrated microfluidics with degassing and pulse-free pumps are required for reliable kinetic constants.
At a glance
- Price range
- Varies by listing
- Category
- Spectroscopy, Analysis & Detection Instruments
- Last updated
- Nov 28, 2026
Current offers
Highlights
- Kretschmann prism coupling remains the dominant architecture; angular interrogation resolves ~10^-6 RIU while intensity interrogation trades resolution for speed and low cost.
- Sensor chips use a ~50 nm gold film on glass; self-assembled alkanethiol monolayers (e.g., 11-mercaptoundecanoic acid) provide carboxyl surfaces for amine coupling of ligands.
- Flow cells range from single-channel to 4-channel or array formats; integrated microfluidics with degassing and pulse-free pumps are required for reliable kinetic constants.
- SPR detects mass changes down to ~1 pg/mm²; low-molecular-weight analytes (<200 Da) need sandwich or competitive formats because direct response scales with mass.
Best use cases
- Best fit: academic or pharma kinetics lab measuring kon/koff/KD on proteins — pick a 4-channel angular-scanning instrument with automated sample loading and temperature control (±0.01 °C).
- Best fit: food/allergen or environmental screening needing multiplexed detection — choose an SPR imaging system with array chips and a fixed-angle CCD, accepting ~10× lower sensitivity than scanning SPR.
- Wrong fit: quantifying small molecules (<200 Da) directly at low concentration — skip direct SPR and use competitive inhibition or switch to ELISA/SPR sandwich assays.
- Wrong fit: high-throughput primary screening of thousands of compounds — skip SPR and use thermal shift or mass spectrometry; SPR is for secondary confirmation of hits.
Buying tips
- Match channel count to throughput: 1–2 channels for single interactions, 4 channels for replicate kinetics, and array/imagine formats only when spatial multiplexing is essential.
- Require a temperature-controlled flow cell (±0.01 °C stability) and automated degassing; unregulated temperature causes >5% baseline drift and invalidates kinetic fits.
- Confirm sensor-chip compatibility: standard gold chips for amine coupling, Ni-NTA or streptavidin chips for His-tagged or biotinylated ligands — buy chips from the instrument vendor to ensure prism index matching.
- For low-analyte-mass work, specify a instrument with noise <0.1 RU and drift <0.1 RU/min; otherwise choose a higher-mass sandwich format or a different technique.
Specs
| Detection Architecture | Prism-coupled angular scan vs imaging |
|---|---|
| Channel Count | 1-2 single / 4-channel / array imaging |
| Sensitivity (RIU) | 10^-6 scanning / 10^-5 imaging |
| Flow Cell Temp Control | ±0.01 °C to ±0.1 °C stability |
Related searches
What to check before buying
- Bench space and services: confirm footprint (typically 60–90 cm width) plus a dedicated PC and a regulated 100–240 V supply with ground.
- Buffer and waste: plan for degassed HBS-EP or PBS running buffer at 0.5–1 mL/min and a waste carboy; insufficient degassing causes bubble artifacts.
- Temperature environment: install in a 20–25 °C room with <2 °C fluctuation and away from direct sunlight or HVAC vents to stabilize baselines.
- Chip and ligand supply: budget for gold sensor chips (reusable 50–100 cycles with regeneration) and coupling reagents (EDC/NHS, ethanolamine) before first run.