Tga Thermal Gravimetric Analysis Buying Guide

A TGA (Thermogravimetric Analyzer) measures mass change versus temperature to quantify composition, thermal stability, and decomposition kinetics. The core trade-off is balance resolution and furnace range versus corrosion resistance and atmosphere control: a 0.1 µg balance with a 1000 °C furnace is compromised if the purge gas or evolved volatiles attack the hang-down wire. Buyers should reject any unit that cannot document balance capacity, temperature calibration method, and gas-flow control, because those three specifications define usable data quality.

Balance Resolution:0.1 µg / 1 µg / 10 µg tiersFurnace Max Temp:1000 °C Kanthal vs 1500 °C Pt-RhAtmosphere Control:Dual MFC N2/O2 / vacuumSample Capacity:100 mg to 5 g load range

Key takeaways

  • Balance resolution tiers: 0.1 µg for micro-samples under 100 mg, 1 µg for standard 200–1000 mg crucibles, and 10 µg for high-capacity 5 g loads; resolution must be paired with a hang-down design that resists buoyancy drift.
  • Furnace and sensor architecture: platinum-rhodium versus Kanthal elements set the upper temperature (1500 °C versus 1000 °C), while thermocouple type (Type S/R vs Type K) determines accuracy above 1000 °C.
  • Atmosphere control separates research-grade from routine units: dual MFC purge lines for N2/O2 switching, vacuum capability, and a water-cooled jacket are required for oxidative stability and evolved-gas coupling.

At a glance

Price range
Varies by listing
Category
Spectroscopy, Analysis & Detection Instruments
Last updated
Nov 28, 2026

Current offers

No live listings right now. Search directly on:

Highlights

  • Balance resolution tiers: 0.1 µg for micro-samples under 100 mg, 1 µg for standard 200–1000 mg crucibles, and 10 µg for high-capacity 5 g loads; resolution must be paired with a hang-down design that resists buoyancy drift.
  • Furnace and sensor architecture: platinum-rhodium versus Kanthal elements set the upper temperature (1500 °C versus 1000 °C), while thermocouple type (Type S/R vs Type K) determines accuracy above 1000 °C.
  • Atmosphere control separates research-grade from routine units: dual MFC purge lines for N2/O2 switching, vacuum capability, and a water-cooled jacket are required for oxidative stability and evolved-gas coupling.
  • Crucible material dictates chemistry: alumina for inert runs to 1500 °C, platinum for oxidizing atmospheres and reactive samples, and aluminum only for sub-600 °C polymer work; wrong crucible material causes catalytic artifacts.

Best use cases

  • Best fit: a 0.1 µg micro-balance TGA with dual MFC gas switching for pharmaceutical polymorph and hydrate stability studies; skip high-capacity 5 g units, which trade resolution for load range you do not need.
  • Best fit: a 1 µg balance, 1000 °C Kanthal-furnace TGA with nitrogen purge for routine polymer filler and moisture-content QC; skip 1500 °C platinum-furnace models unless your method requires ceramic or metal oxidation data.
  • Best fit: a 10 µg, 5 g capacity TGA with a large crucible for heterogeneous industrial samples like coal, cement, or ores; skip micro-balance units, which cannot handle representative sample masses.
  • Wrong fit: buying a TGA without mass-spectrometer or FTIR coupling ports when your goal is evolved-gas identification; select a unit with a heated transfer line interface instead of adding aftermarket adapters.

Buying tips

  • Match balance capacity to sample mass: a 200 mg capacity cannot run 500 mg heterogeneous samples, and a 5 g load cell typically resolves only 10 µg, so define your required resolution before choosing capacity.
  • Require a temperature calibration kit or certified reference materials (Curie-point standards such as alumel, nickel, perkalloy) and confirm the furnace linearity specification across the full range, not just at one point.
  • Specify purge-gas mass flow controllers (MFCs) with at least two calibrated lines for N2 and O2; manual needle valves cause baseline drift and make oxidative induction time (OIT) results non-comparable.
  • Check the hang-down wire and thermocouple materials against your sample chemistry: platinum hardware for oxidizing or halogen-containing samples, quartz or alumina for inert runs, and avoid bare Ni-Cr wire with sulfur-bearing samples.

Specs

Balance Resolution0.1 µg / 1 µg / 10 µg tiers
Furnace Max Temp1000 °C Kanthal vs 1500 °C Pt-Rh
Atmosphere ControlDual MFC N2/O2 / vacuum
Sample Capacity100 mg to 5 g load range

Related searches

What to check before buying

  • Verify lab bench load capacity and clearance: TGA furnaces require 30–50 cm of vertical clearance above the balance for the furnace lift and hang-down assembly.
  • Confirm facility gas supply: dual regulated purge lines (N2 and O2) at 99.999% purity with two-stage regulators, plus a dedicated exhaust for evolved decomposition products.
  • Check electrical service: most 1500 °C TGAs require 208–240 V single-phase or three-phase service and a stable grounding point; confirm the plug standard and breaker rating before delivery.
  • Measure crucible and sample budget: confirm the balance capacity in grams, the standard crucible volume (typically 70–150 µL), and whether an autosampler carousel fits your sample throughput.